Parameters of porcelain oven |
|
Maximum Temperature |
1100℃ |
The highest vacuum |
-98Kpa |
Temperature control accuracy |
±1℃ |
Heating rate |
≤140℃/min |
Power rating |
1500W |
Input voltage |
220/110V 50/60HZ |
Furnace size |
∅120*70mm |
The number of program |
100 Articles |
Outline dimension |
Length * width * height =380 * 299 * 565mm |
Weight of equipment |
30kg |
Troubleshooting |
failure cause |
The exclusion method |
The heating system is abnormal |
The actual temperature of the furnace exceeds the maximum value |
Check that the thermocouple is disconnected Check whether the SCR is broken down. |
Abnormal vacuum system |
The vacuum degree did not meet the requirement in 60s. |
Check the vacuum line for reliable connection or cracking. Check that the vacuum pump is running. |
Roast Furnace Structure
● Display screen
● Trays
● The burning table
● Heating furnace
● Roast furnace main body
● The tray was temporarily placed with the tooth crown
● Mains switch
● 250V 3A insurance pipe
● Power input socket (with 250V 8A insurance tube)
● Vacuum pump power supply socket
● Vacuum pipe interface
● Airway